Traceable thermometer/alarm
Every MED unit is equipped with a Traceable
Thermometer/Alarm that simultaneously displays
minimum, maximum and current interior
temperatures and provides a visual and audible
signal when the temperature inside the unit falls
outside the set temperature range.
(MED models):
The temperature sensor, or probe, is enclosed in
a glycol-filled bottle that you insert into the
location to be monitored, such as inside a
refrigerator or freezer, once the unit has been
programmed. The thermometer/alarm itself can
be attached to the door of the unit with the
magnet on the back, or it can be placed on a
convenient table or shelf.
Power to the Traceable Thermometer/Alarm is supplied with a lithium iron disulfide
battery with a 10-year shelf life. When it is inserted correctly, the unit will beep and all
segments of the LCD display will be temporarily activated. (Note that alarm
temperatures will have to be reset after replacing the battery.)
The temperature display can be switched between Celsius (°C) and Fahrenheit (°F)
by sliding the
°C/°F button located on the back of the unit.
Function of Controls:
MODE/RESET Key: Press ONCE to reset Max/Min temperature record. When the unit
is sounding an alarm, press ONCE to turn off the beeping sound.
HIGH/LOW Key: Press ONCE to display high alarm setting. Press TWICE to display
low alarm setting. Press THREE TIMES to display current temperature reading.
ALERT ON/OFF Key
Programming the Alarms:
: When the unit is sounding an alarm, press ONCE to turn off.
However, the High-Alarm or Low-Alarm will still flash until the temperature goes
back to normal or the Alarm function is turned off.
The thermometer displays the internal temperature of the unit being monitored, as
well as the maximum and minimum temperatures reached. The maximum and
minimum readings reflect the highest and lowest temperatures since the battery was
inserted or since the MODE/RESET key was last pressed. If the temperature rises
above or falls below the set temperature range, the alarm will sound. By default, the
high alarm temperature setting is 8°C (46.4°F) and the low alarm temperature setting
is 2°C (35.6°F). To reset these limits, follow these instructions:
High Alarm Temperature Setting: Press the HIGH/LOW key ONCE, then press
MODE/RESET to enter the High Alarm temperature setting mode. Use the Up and
Down arrows to adjust the temperature accordingly. Press MODE/RESET once
again to confirm the setting.
Low Alarm Temperature Setting: Press the HIGH/LOW key TWICE, then press
MODE/RESET to enter the Low Alarm temperature setting mode. Use the Up and
Down arrows to adjust the temperature accordingly. Press MODE/RESET ONCE to
confirm the setting.